drmscorpus · anaesthesia physics

What the Breath
Tells You.

Capnography and capnometry — and why infrared beat Raman.

The monitor says SpO₂ 99%.
The heart rate is steady, the patient looks pink, and everything seems fine. But a pulse oximeter tells you about oxygenation, not ventilation. A patient breathing supplemental oxygen can hold a normal SpO₂ while CO₂ quietly climbs.

Capnography was built to fill that gap. This post covers what it measures, how to read the trace, what it is used for, where it misleads, and why infrared absorption beat Raman scattering.

New here? The pulse oximeter post covers that blind spot first.
scroll
§ 01

Oxygenation is not Ventilation

The two monitors answer different questions

The two monitors answer different questions, and neither replaces the other.

Pulse oximetryCapnography
MeasuresSpO₂, haemoglobin saturationCO₂ in exhaled gas, breath by breath
Tells you aboutOxygenationVentilation, perfusion and metabolism
PrincipleBeer Lambert’s lawBeer Lambert’s Law
In apnoea on 100% oxygenFalls late, often after minutesTrace goes flat immediately
Blind spotPaCO₂, hyperoxiaPaO₂, oxygen delivery
§ 02

Capnometry or Capnography?

A number, or a number and a picture

Capnometry is the measurement of CO₂ in the airway gas, shown as a number: the end-tidal CO₂ (ETCO₂). Capnography is the continuous graph of CO₂ against time, called the capnogram. A capnometer gives you a number. A capnograph gives you a number and a picture, and most of the information is in the picture.

Normal ETCO₂
35–45
mmHg

Normal end-tidal CO₂

4.7–6.0
kPa

The same range in kilopascals

2–5
mmHg gap

ETCO₂ sits below the arterial value in a healthy lung

The gap exists because some alveoli are ventilated but not perfused, and they dilute the expired gas.

§ 03

Where the CO₂ Comes From

Production, perfusion and ventilation

CO₂ is made by cells, carried in the blood (mostly as bicarbonate), delivered to the lungs by the heart, and removed by ventilation. ETCO₂ therefore depends on three things at once: production, perfusion and ventilation. A change in any one moves the number.

ChangeETCO₂Examples
Production upRisesFever, shivering, sepsis, malignant hyperthermia
Production downFallsHypothermia, deep anaesthesia
Perfusion downFallsLow cardiac output, pulmonary embolism, cardiac arrest
Ventilation downRisesHypoventilation, rebreathing, airway obstruction
Ventilation upFallsHyperventilation
§ 04

Reading the Capnogram

Each breath draws one waveform

Each breath draws one waveform, and each part has a meaning.

PhaseSegmentWhat is happening
0Inspiratory downstrokeFresh gas passes the sensor, so CO₂ falls to about zero
IExpiratory baselineDead-space gas with no CO₂
IIExpiratory upstrokeDead-space gas mixes with alveolar gas, giving a steep rise
IIIAlveolar plateauAlveolar gas. The ETCO₂ is the highest point at the end of this phase
Normal capnogram showing phases 0, I, II and III, the alpha and beta angles and the ETCO₂ point

Fig 1 — The normal capnogram: phases and angles

Two angles worth a glance

The α angle (between phases II and III) widens when expiration is obstructed. The β angle (between the plateau and the inspiratory downstroke) widens with rebreathing.

§ 05

What the Shape Tells You

Read the waveform first, the number second

FindingLikely cause
Sudden drop to zeroDisconnection, oesophageal intubation, blocked tube, cardiac arrest
Sudden fall, not to zeroPulmonary or air embolism, severe hypotension, partial leak
Gradual fallHyperventilation, falling cardiac output, hypothermia
Gradual riseHypoventilation, rising temperature, malignant hyperthermia
Raised baselineRebreathing, for example exhausted soda lime or a faulty expiratory valve
Sloping, “shark-fin” plateauBronchospasm, kinked tube, obstructive lung disease
Dip in the plateau (curare cleft)Return of spontaneous effort in a partly paralysed patient
§ 06

What It Is Used For

From intubation to CPR

§ 07

Mainstream or Sidestream?

Where the sensor sits

MainstreamSidestream
Sensor positionIn an adapter at the airwayIn the monitor, with gas drawn through a tube (about 50–300 mL/min)
ResponseFastSlightly delayed by the sampling line
At the airwayAdds weight and a little dead spaceLight
MoistureHeated window limits condensationNeeds a water trap, and tubing can clog
Typical useIntubated patientsIntubated and non-intubated patients
§ 08

The Physics: Four Things Light Can Do

Reflection, transmission, absorption, scattering

When light meets matter, the incident energy is split four ways:

Conservation of light
Incident = reflected + transmitted + absorbed + scattered

Infrared capnometry relies on absorption. Raman spectroscopy relies on scattering. Everything else follows from that difference.

§ 09

Infrared Absorption: How Most Capnometers Work

CO₂ absorbs at about 4.26 µm

CO₂ absorbs infrared radiation strongly at about 4.26 µm. A capnometer passes infrared light through the gas sample, and a detector measures how much arrives at the other side. The more CO₂ present, the less light gets through. The Beer-Lambert law links the two, so the monitor can convert transmitted light into a CO₂ concentration or partial pressure.

The common design is non-dispersive infrared (NDIR). A broadband source and an optical filter select the CO₂ wavelength band, so no prism or grating is needed. That keeps the device small, cheap and robust.

Two practical points that come up in exams

Nitrous oxide overlaps. N₂O absorbs near 4.5 µm, close to the CO₂ band, and can broaden the absorption. Monitors correct for this.

Homonuclear molecules are invisible. O₂ and N₂ have no change in dipole moment when they vibrate, so they do not absorb infrared. IR cannot measure them.

§ 10

Raman Scattering: The Fingerprint Approach

Inelastic scattering, one laser, every gas

Flowchart: when a photon meets a molecule, elastic Rayleigh scattering versus inelastic Raman scattering (Stokes and anti-Stokes)

Fig 2 — When a photon meets a molecule

When monochromatic laser light hits a gas molecule, most of it is scattered with no change in energy. This is Rayleigh scattering, and it is elastic. A very small fraction is scattered with a shifted energy. This is Raman scattering, and it is inelastic.

Energy-level diagrams for Rayleigh, Stokes and anti-Stokes scattering

Fig 3 — Energy levels: Rayleigh, Stokes and anti-Stokes

The size of the shift depends on the molecule’s vibrational energy levels, so each gas produces its own set of peaks. A single laser and one detector can therefore identify and quantify O₂, N₂, CO₂, N₂O and the volatile agents together, including the gases IR cannot see. Stokes lines are usually stronger at room temperature, because most molecules start in the lowest vibrational level.

Raman gas analysers were used clinically, for example in multi-gas systems such as the Ohmeda RASCAL.

§ 11

Why Raman Did Not Win

The problem is signal strength

Raman scattering is extremely weak, roughly a millionth of the Rayleigh signal or less. Detecting it needs a laser, precise optics and a sensitive detector. That made the monitors bulkier, more costly and less rugged than IR units.

Side by sideInfrared absorptionRaman scattering
Physical principleAbsorptionInelastic scattering
Gases measuredCO₂, N₂O, volatile agentsO₂, N₂, CO₂, N₂O, volatile agents
Sees O₂ and N₂?NoYes
Signal strengthStrongVery weak
Size and costSmall, cheapLarger, costlier
RobustnessHighLower
Current useStandard in clinical practiceLargely historical
§ 12

So Which Is Ideal?

It depends on how the question is asked

Viva tip

State both and then commit: IR absorption wins in practice because the CO₂ signal is strong and specific, and a cheap, small, reliable sensor matters more than the ability to measure every gas.

Other methods worth a mention:

§ 13

When Capnography Misleads

A signal, not a verdict

Like the pulse oximeter, capnography is a signal, not a verdict. Read the waveform first and the number second.

LimitationEffectWhy
ETCO₂ is not PaCO₂ETCO₂ reads lower than arterialAlveolar dead space from pulmonary embolism, low cardiac output or COPD widens the gap
Low cardiac output or arrestLow or zero readingLittle CO₂ reaches the lungs, so the monitor is correct and the patient is not
Sample line problemsLow or distorted traceKinks, water or leaks dilute or block the sample
Non-intubated patientsUnderestimates ETCO₂Room air and oxygen dilute the expired gas
Small patients, fast ratesUnderestimates ETCO₂Sidestream sampling flow and slow rise time blunt the trace
Nitrous oxide and high oxygenReading errorCollision broadening alters CO₂ absorption unless the monitor compensates
CondensationDropout or false valuesWater droplets absorb infrared
Colorimetric detectorsFalse positive or negativeCarbonated drinks or gastric acid can change colour, and low flow can fail to
Tube positionNormal trace despite wrong depthCapnography confirms the trachea, not an endobronchial tube
§ 14

A Short Relay

1928 to the 1990s

§ 15

Key Points

For your next viva

  1. The pulse oximeter measures oxygenation. Capnography measures ventilation, perfusion and metabolism.
  2. Capnometry gives the number. Capnography gives the waveform, and the waveform holds most of the information.
  3. IR capnometry uses absorption of IR light at about 4.26 µm.
  4. Raman uses inelastic scattering and gives a fingerprint for each gas, including O₂ and N₂.
  5. Raman’s weak signal made it costly and bulky, so IR remains the standard.
  6. ETCO₂ is not PaCO₂, and a flat trace needs action, not assumptions.
§ 16

Full Circle

A flat line after intubation

The Scenario
You have just intubated. SpO₂ reads 99% on 100% oxygen, and the chest seems to move. But the capnogram is flat.

What do you do? What is your first thought, and what do you check next? Drop your approach in the comments below. It is not only a question about capnography. It is a question about how you think at the bedside.

अपाने जुह्वति प्राणं प्राणेऽपानं तथापरे।
प्राणापानगती रुद्ध्वा प्राणायामपरायणाः॥ ४.२९॥
Some offer the incoming breath into the outgoing, and the outgoing into the incoming, devoted to the discipline of the breath.
Bhagavad Gita 4.29

For centuries, people have paid attention to the breath. Capnography is the modern way of listening to it, one waveform at a time.